Refurbished Mettler Toledo XP26DR Micro Balance

Refurbished Mettler Toledo XPE205DR/M microbalance is ideal for weighing small samples directly into the tare container, as they allow very low minimum sample weights.

Refurbished Mettler Toledo XP26DR Micro Balance

Refurbished Mettler Toledo XPE205DR/M microbalance is ideal for weighing small samples directly into the tare container, as they allow very low minimum sample weights.  List price Refurbished Mettler Toledo XP26DR Micro Balance when new approx £20k

Standard features

  • SmartSens, Sensor for hands-free operation
  • SmartGrid, grid weighing pan for fast and stable results
  • SmartScreen, colour display for safe and user-friendly operation
  • Holder for tare containers for ergonomic weighing
  • MinWeigh warns if the minimum weight is not reached (Minimumweight determined on-site by a service technician)
  • User Management for granting access rights
  • LevelControl warns when the balance is not levelled
  • BalanceCheck, prompts automatically for validation with external weights
  • ProFACT, fully-automatic temperature and/or time-controlled internal adjustment and linearisation
  • Seven different applications with user guidance (incl. density determination, statistical analysis, formulation and differential weighing)
  • Individual reports can be defined with up to 4 IDs
  • Built-in RS232 interface and two auxiliary outlets
  • Slot for second interface (7 options)
  • Height-adjustable inner draft shield
  • Motor-driven draft shield doors
  • Completely dismountable draft shield for easy cleaning
  • Terminal can be positioned separately from the balance
  • Protective cover for the terminal

Technical data (Limit Values)XP26DR
Maximum capacity22 g
Maximum capacity, fine range5.1 g
Readability0.01 mg
Readability, fine range0.002 mg
Repeatability
– (at nominal load)0.008 mg (20 g)
– (at low load)0.005 mg (1 g)
– (at low load), fine range0.002 mg (1 g)
Linearity0.01 mg
Excentric load deviation (test load)0.025 mg (10 g)
Sensitivity offset5×10–6·Rnt
Sensitivity temperature drift21×10–6/°C·Rnt
Sensitivity stability 3)1×10–6/a·Rnt
Interface update rate23/s
Typical values for calculating the measurement uncertainty
Repeatability* (sd)0.004 mg+5×10–8·Rgr
Repeatability*, fine range (sd)0.0012 mg+1×10–7·Rgr
Differential linearity deviation3×10–13 g·Rnt
Differential eccentric load deviation (sd)4×10–7·Rnt
Sensitivity offset (sd)2)1.2×10–6·Rnt
Minimum weight (according to USP3.6 mg+3×10–4·Rgr
Minimum weight * (U=1%, 20.24 mg+2×10–5·Rgr
Settling time2.5s
Settling time, fine range3.5s

Additional information

Capacity

Readability